Technical Specifications

Assay Types / Modes 8 modes (Absorbance, End Point, Fixed Time, Rate, S ample Blank, Multi Standard, Two point Non Linear, and Sample Blank Non Linear).
Incubator 4 Incubator Positions
Sample Type Serum / Urine / Plasma / Body Fluids / Hemolysate
Light Source 6 V, 10 W Quartz Halogen Lamp (Life Span approx 2000 Hours)
Measurement Method Monochromatic and Bichromatic
Photodiodes One Silicon photodiode
Wavelength (in Filter wheel) 7 Filters (340, 405, 505, 546, 578, 620, 670 nm) and 1 Free position
Absorbance Range 0 - 3 Absorbance
Resolution 0.0001
Stray light ≥ 2.3 Absorbance
Delay/Incubation Time 3 - 999 seconds
Read Time 3 - 999 seconds for end point and 30 - 999 for rate and fixed time.
Types of Test ware Quartz Flowcell, 6 mm polystyrene cuvette
Flowcell Volume 32 μl
Display 240 x 128 pixel blue color graphical LCD with LED backlight.
Keyboard Yes
Printer In built 384 dots Thermal Printer
PC Communication 9 pin D type Serial Connector
PC Software Optional
Memory FLASH
Storage 25,600 Test Data Storage, 3 Month QC Data
Result Reports Collated by ID, Date and Test
Quality Controls Daily and Monthly for 3 levels with LJ Graph
Aspiration Volume 200 - 1500 μl
Air Aspiration Volume Variable (To reduce Carry over & Cross contamination)
Cross Contamination <1%
Temperature Control 25°C, 30°C, 37 °C Accuracy ±0.1°C, Peltier control
Operating Voltage 230 V, 50 Hz, 45 Watt max
Storage Temperature -10 to 40 °C
Operating Temperature 10 to 35 °C
Humidity ≤ 80% RH
Dimensions 250 mm(W) x 290 mm(D) x 415 mm(H), Weight ~ 11 kgs